Which type of gas should i pick for a purified glove box?

The glove box should be supplied with inert gas with a 2 bars pressure. Normally, the inert gas chosen is nitrogen or argon and in some cases helium.

Regeneration requires a bottle composed either of hydrogen and inert gas or inert gas alone according to the glove box’ purification type;

Input

Main feeding

Reactor’s regeneration gas

O2 Purification

O2 and H2O Purification 

H2O Purification
Gas

Inert gas (Ar, N2,He)

95 % Inert gas
5 % H2
100%
Inert gas

Pressure/Flow

2 bars pressure

Flow = 23 L.mn-1

Type of tank

Tank or Inert gas
network quality :
[O2] < 5 ppm ; [H2O] < 5 ppm

10 m3 Tank

Connection

Permanent

Punctual
(only during the reactor’s regeneration phase)
Fitting

4-6 or F1/4″ BSP flexible hose

4-6 or F1/4″ BSP flexible hose